Chapter 6 Operations and Measurements
6. -Width: indicates the time between the threshold middle value of a falling edge
to the threshold middle value of the next rising edge.
7. +Duty: indicates the ratio of the positive pulse width to the period.
8. -Duty:indicates the ratio of the negative pulse width to the period.
9. Tvmax: indicates the time that corresponds to the maximum value of the
waveform (Vmax).
10. Tvmin: indicates the time that corresponds to the minimum value of the
waveform (Vmin).
Note:
The default values for threshold upper limit, threshold middle value, and
threshold low limit are 90%, 50%, and 10%, respectively. You can press Measure
→ Setting to enter the sub-menu to set the parameters.
Count Values
1. Positive Pulse Count: The number of positive pulses that rise from under the
threshold lower limit to above the threshold upper limit.
2. Negative Pulse Count: The number of negative pulses that fall from above
the threshold upper limit to below the threshold lower limit.
3. Rising Edge Count: The number of rising edges that rise from under the
threshold lower limit to above the threshold upper limit.
MSO5000-E User Guide
1
2
1
2
Positive Pulse Count = n
n
Threshold Upper Limit
Threshold Lower Limit
Negative Pulse Count = n
Threshold Upper Limit
Threshold Lower Limit
n
RIGOL
6-31