17B203=* E-MAG Active
NOTICE:
The cable between the orthotic knee joint and elec
tronics may not be shortened or extended.
INFORMATION: If there is excess cable length, lay
the cable in flat loops and secure it in the casing
with double-sided adhesive tape.
To install the electrical components, remove the dum
mies for the electronics and battery from the recept
acles.
Connect the two cables to the electronics receptacle.
Ensure the cable clips engage properly.
INFORMATION: Leave approximately 5 cm of addi
tional cable length to allow for installation of the
locking unit.
Optionally, you may shorten the battery cable in the bat
tery plug section. Ensure that the cable cross-section is
maintained.
INFORMATION: Ensure the correct polarity is main
tained after shortening: plus cable (red) to plus
connection (+ symbol on locking unit), minus cable
(grey) to minus (no symbol).
Connect the battery plug to the locking unit.
Preparation for use