NOTICE:
The cable between the orthotic knee joint and elec­
tronics may not be shortened or extended.
INFORMATION: If there is excess cable length, lay
the cable in flat loops and secure it in the casing
with double-sided adhesive tape.
To install the electrical components, remove the dum­
mies for the electronics and battery from the recept­
acles.
Connect the two cables to the electronics receptacle.
Ensure the cable clips engage properly.
INFORMATION: Leave approximately 5 cm of addi­
tional cable length to allow for installation of the
locking unit.
Optionally, you may shorten the battery cable in the bat­
tery plug section. Ensure that the cable cross-section is
maintained.
INFORMATION: Ensure the correct polarity is main­
tained after shortening: plus cable (red) to plus
connection (+ symbol on locking unit), minus cable
(grey) to minus (no symbol).
Connect the battery plug to the locking unit.
Preparation for use