Otto Bock E-MAG Active 17B203-L Instructions For Use Manual page 32

Hide thumbs Also See for E-MAG Active 17B203-L:
Table of Contents

Advertisement

Available languages

Available languages

Preparation for use
6
7
8
CAUTION
Risk of injury due to joint bars breaking
A broken joint bar can result in injury to the patient. Carefully sand down and smooth fissures and ridges on the
lamination bars before laminating the orthosis.
NOTICE
Risk of damage due to improper installation
When setting up the orthosis, ensure that the bending radius is not less than 15 mm to avoid the risk of breaking.
Grind away surface imperfections such as grooves and burrs caused during shaping. The joint bars must not be
subjected to heat treatment.
INFORMATION
The sole function of the casing is to protect and secure the battery, electronics and cable connections. The cas­
ing only has to be large enough to enclose these components.
Remove all lamination dummies except for the E-MAG knee joint dummy and set up the lamination bars. The drilled
holes in the insertion area of the knee joint are used to temporarily fasten the lamination bars with a screw during
the alignment of the orthosis until the fitting.
32
Select the soft tissue regions and adapt the fibre mesh
to the shape of the lamination dummies. Adapt the
plaster positive in the selected regions so the lamination
dummies lie flat.
Note the length of the cables when positioning the com­
ponents:
Battery – electronics: 500 mm, variable length
Electronics – knee joint: 300 mm, fixed length
Mark the final position of the lamination dummies onto
the region of the thigh:
30Y103 lamination dummy 3 for battery 2 with
receptacle 1.
30Y100 lamination dummy 3 for electronics 2 with
receptacle 1.
17B203=* E-MAG Active

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents