Controls and Indicators
2.1.9 Patient Current Indicator <9>
If the indicator does not flash during treatment, you have to carry out a self-test routine
and observe whether or not the indicator flashes. If the indicator still does not flash during
treatment after the check, the accessories should be checked and replaced if necessary.
2.1.10 Output Indicator <10>
2.1.11 Patient Lead Connector <11>
11
The patient lead connector <11> serves to plug in the patient lead. You can plug plate
electrodes, adhesive electrodes or other types of electrodes into the patient lead.
Patient Lead - Mode(s) of Attachment
The colours of the connectors facilitate easy and correct
attachment of the electrodes according to their polarity.
Colour of the female plug:
red = anode (+), blue = cathode (-)
Type BF component, not connected to
protective ground wire.
Warning Take care that the electrodes are by no means touched after the current is
turned up!
6
The flashing of the indicators depends on the resistance
in the circuit. In most cases the diodes start flashing ex-
actly at the moment when the patient starts to have a clear
sensation of the current.
The indicator tells you to be cautious when handling the
electrodes:
The patient lead connector <11> is under
Warning!
voltage! Take care that the electrodes are by no means
touched after the current is turned up!
12
Anode (+)
Cathode (-)