2.4 Characterisation of the Individual Current Modes

All current modes of the unit are emitted in the
Note
the exception of simultaneous treatment (see page 37). Here, the device is
automatically switched to a mixed mode consisting of
constant voltage mode
tact between skin and transducer is interrupted.

2.4.1 Low-frequency Current Modes (LF)

(G)
Galvanisation
Direct current without any interruption or break
Variation:
Medium-frequency interrupted direct current (8 kHz; duty cycle 95%).
Application:
Note
You have to take special care when applying galvanic current. The mois-
tened viscose covers must be well attached to the skin surface. The intensity
must not exceed 0.1 mA/cm² of active electrode surface!
(DF)
Diadynamic Current
acc. to Bernard
Fixed parameters
Pulse shape: sinusoidal half-waves
T = pulse time: 10 ms
R = release time: 0 ms
Frequency: 100 Hz
Galvanic basis ON
PHYSIOMED-Jubilee
to avoid unpleasant current sensations when the con-
Iontophoresis, basic therapy for paralysis and atrophy treatment,
evoking hyperaemia
Diphasé Fix e
OFF
constant current mode
Adjustable
Galvanic basis (5%)
Controls and Indicators
, with
constant current
and
13