Error Messages - Hartmann Veroval BP750X Instructions For Use Manual

Ecg and blood pressure monitor
Table of Contents

Advertisement

Available languages

Available languages

English
EN
Start the program and connect the Veroval
to your PC using the enclosed USB Bluetooth
USB Bluetooth
®
ECG and blood pressure monitor transfers the saved measured values via
Bluetooth
directly to the Veroval
®
instructions in the Veroval
No data should be transferred during a measurement.
is displayed on the ECG and blood pressure monitor.
Start the data transfer using the 'medi.connect' software.
During the data transfer, the data transfer symbol
The ECG and blood pressure monitor switches off automatically after 30 seconds
of inactivity and when communication with the PC is interrupted.

10. Error messages

Error that has
occurred
Monitor cannot
be switched on
Cuff will not
inflate
The pulse value
flashes when the
ECG measure-
ment is displayed
EE
adapter into the USB port of your computer. The Veroval
®
medi.connect software.
®
Possible causes
Batteries are missing,
inserted incorrectly or
flat.
Cuff connector plug is
incorrectly positioned in
the connector socket of
the monitor.
Wrong cuff type con-
nected.
Pulse could not be
detected correctly.
Invalid ECG measurement Check whether there is sufficient
ECG and blood pressure monitor
®
adapter. To do this, insert the
®
medi.connect software. Then follow the
Remedy
Check batteries and insert
four identical new batteries if
necessary.
Check the connection between
the cuff connector plug and
socket.
Verify that only an approved
Veroval
®
being used.
Repeat the ECG measurement.
skin contact to the upper and
lower electrodes when recording
the ECG.
62
®
flashes.
cuff and connector are