8. Error messages

Error which has
occurred
Monitor cannot be
switched on
E2
E3
E4
E5
E6
Possible causes
Batteries are missing, inserted incorrectly or flat.
The measuring signals were not read or were read
incorrectly. This can be caused by the cuff not
being applied correctly, moving, talking or a very
weak pulse.
Invalid measurement due to movement.
The cuff was not applied correctly.
Error during the measurement.
Inflation pressure is higher than 300 mmHg.
There is a system error.
Remedy
Check batteries and insert two identical new batter-
ies if necessary.
Check that the cuff has been positioned correctly.
Do not talk or move during the measuring proce-
dure. Also observe the 10 golden rules.
Do not talk or move during the measuring proce-
dure.
Check that the cuff has been positioned correctly.
Contact customer service if this error message
appears.
Please rest for at least 1 minute and take the meas-
urement again.
Contact customer service if this error message
appears.