Error that has
occurred
E 0
E 2
Possible causes
No pulse could be
detected during the
blood pressure mea-
surement because the
cuff was not correctly
applied.
The cuff pressure is grea-
ter than 300 mmHg.
No pressure could be
built up in the cuff.
Unusually high or low
blood pressure measured
values were detected.
Remedy
Check that the cuff has been
positioned correctly. Apply the
cuff so that two fingers fit bet-
ween the cuff and the upper
arm.
Air tube is not correctly con-
nected to the monitor. Check
that the connector plug is cor-
rectly positioned. If this error
occurs frequently, you should
use a new cuff.
Check that the cuff has been
positioned correctly. Apply the
cuff so that two fingers fit bet-
ween the cuff and the upper
arm.
Air tube is not correctly con-
nected to the monitor. Check
that the connector plug is cor-
rectly positioned. If this error
occurs frequently, you should
use a new cuff.
Check that the cuff has been
positioned correctly. Apply the
cuff so that two fingers fit bet-
ween the cuff and the upper
arm.
Air tube is not correctly con-
nected to the monitor. Check
that the connector plug is cor-
rectly positioned. If this error
occurs frequently, you should
use a new cuff.