Available languages

7. Explanation of error messages

Error that has
Possible causes
occurred
Monitor cannot be
Batteries are missing, inserted incorrectly or flat.
switched on
Cuff will not inflate
Cuff connector plug is incorrectly positioned in
the connector socket of the monitor.
Wrong cuff type connected.
E  I
Pulse could not be detected correctly.
E2
The pulse could not be detected.
E3
The cuff was applied too firmly or too loosely.
Remedy
Check batteries and insert four identical new batteries if
necessary.
Check the connection between the cuff connector plug and
socket.
Verify that only an approved Veroval
being used.
Check whether the cuff has been applied correctly. Do not
talk or move during the measuring procedure.
Check whether the cuff has been positioned correctly. Do not
talk or move during the measuring procedure.
Apply the cuff so that two fingers fit between the cuff and
the upper arm.
Air tube is not correctly connected to the monitor.
Check that the connector plug is correctly positioned.
If this error occurs frequently, you should use a new cuff.