Revision B
Maintenance: Disassembly Guidelines
3. Insert rear bezel into slot on back of MAC 5000 assembly as shown
below.
4. Rotate bezel to the upright position as shown below.
5. With the new bezel in place, replace the top cover by reversing the
steps described previously.
6. Replace the battery and proceed with software, serial number, and
system setups as described in the following sections.
MAC™ 5000 resting ECG analysis system
2024917-010
4-15