Compatible GE Medical Systems Information Technologies Treadmills

Analog Treadmills

3-16
Installation: MAC 5000 ST Requirements and Configuration
MAC 5000 Configuration Requirements – At the Main Menu complete
the following in the order shown below:
Select System Setup,
Enter System password,
Exercise Test,
Inputs/Outputs,
Change Blood Pressure to Suntech.
Ergoline - Model Ergoline 900
Connection Requirements – Use cable PN 2008110-001 to connect
from the MAC 5000 port 1 to the Ergoline serial port. Use cable
PN 2008115-001 to connect from the MAC5000 ANA/TTL port to
the Ergoline QRS trigger input.
Device Configuration Requirements – See Ergoline 900 Operator's
Manual.
MAC 5000 Configuration Requirements – At the Main Menu complete
the following in the order shown below:
Select System Setup,
Enter System password,
Exercise Test,
Inputs/Outputs,
Change Blood Pressure to Ergoline Ergometer.
Model T2000
Connection Requirements – Use cable PN 2007918-001 (T2000) to
connect from the MAC 5000 port 1 to the treadmill serial port.
Device Configuration Requirements – None.
MAC 5000 Configuration Requirements – Use the Edit Protocol
application to set the protocol Test Type to Treadmill in MPH or
Treadmill in Km/H for protocols that will be used with this treadmill.
Connection Requirements – There are no cables available from GE
Medical Systems Information Technologies to interface to analog
treadmills. The customer is responsible for making the appropriate
cable. Speed and grade signals for controlling analog treadmills are
available on pins 2 (Slow Analog Output) and 8 (Fast Analog Output)
of the ANA/TTL port. Pins 1, 4 and 5 are tied to ground.
Device Configuration Requirements – None.
MAC5000 Configuration Requirements – Use the Edit Protocol
application to set the protocol Test Type to Analog Treadmill in MPH
or Analog Treadmill in Km/H for protocols that will be used with this
treadmill.
MAC™ 5000 resting ECG analysis system
2024917-010
Revision B