Advertisement

Software

4-16
Maintenance: Disassembly Guidelines
After replacing the -006 board, you need to install or update the software
on the board as follows:
NOTE
Connect the system to AC power before you begin the software
update. Keep the system connected to AC power during the software
update and do not power off the system during the software update.
1. Press Power to turn on the system.
2. From the Main Menu, select System Setup.
3. Enter the system password. and press Enter.
4. Press Shift + F3.
The message below is displayed.
5. Insert the secure digital card.
A message similar to the one shown below is displayed.
Press 'Enter' to start installation
6. Press the Enter key.
If the system is not connected to AC power, the message shown below
is displayed.
If the message shown above appears on the screen, connect the
system to AC power and continue with step 7.
MAC™ 5000 resting ECG analysis system
2024917-010
Please Insert SD Card
Press 'Esc' to cancel
Current Version:
New Software Version:
Please switch on AC Power !
Press 'Esc' to cancel
Revision B