Manual Temperature Entry; Measurement With Automatic Temperature Compensation - JUMO 202550 Operating Instructions Manual

Up indicator/controller for analytical measurement
Table of Contents

Advertisement

18 Conductivity indicator
Procedure

18.4 Manual temperature entry

Initial condition
Procedure

18.5 Measurement with automatic temperature compensation

Initial condition
Procedure
50
Chapter 21.2 "Analog inputs - C111", page 61.
The instrument is in the measurement mode,
"Operating modes and states", page 24.
The upper display shows the compensated conductivity value of the solution
being measured.
The indicated conductivity depends on the manually set temperature, see
Temperature setting, below
and the set (or automatically acquired) temperature coefficient (TC),
Chapter 18.8.1 "Automatic determination of the temperature coefficient,
using manual temperature entry", page 55.
The lower display shows the manually entered temperature setting.
The temperature acquisition is configured as
"Manual temperature compensation",
Chapter 21.2 "Analog inputs - C111", page 61.
The operating level is unlocked,
"Unlocking the levels", page 27, (code word 0110)
The instrument is in the measurement mode,
"Operating modes and states", page 24.
Press
briefly and repeatedly, until "InP2" is displayed.
PGM
Use the
and
Press the
key (confirmation)
PGM
Press the
key (to return to measurement mode) or cancel the entry
EXIT
The temperature acquisition has been configured as
"Automatic temperature compensation with Pt100 or Pt1000",
Chapter 21.2 "Analog inputs - C111", page 61.
The instrument is in the measurement mode,
"Operating modes and states", page 24.
The temperature measurement for the medium cannot be altered manually.
keys to set the temperature that is shown