Process Value Output For Conductivity - C311; Rang - Cell - Alph - JUMO 202550 Operating Instructions Manual

Up indicator/controller for analytical measurement
Table of Contents

Advertisement

21 Configuration level of the conductivity indicator

21.3 Process value output for conductivity - C311

Bilinear characteristic
0%
1%
...
99%

21.4 rAnG - CELL - ALPH

rAnG
CELL
62
*The factory-set parameters are shown in the position boxes.
The range number is used to select the display range.
Display range
0 — 9999
0.0 — 999.9
0.00 — 99.99
0.000 — 9.999
The rAnG settings 22 to 25 are designed for connecting the
indicator/controller to a 2-wire transmitter that does not have or
cannot use its own calibration routine.
If a transmitter is connected that does have it own calibration
routine for the attached probe, then the rAnG settings 27 to 30
must be used.
The relative cell constant K
of the cell constant from the nominal value (0.01; 0.1; 1.0; 3.0; 10.0) over the
range from 80 to 120%.
Application
Conductivity – no decimal places
Conductivity – one decimal place
Conductivity – two decimal places
Conductivity – three decimal places
[%] can be used to compensate for the deviation
rel
C311* 5
0
I
I
0
0
0
1
9
9
I
I
Range
(rAng)
22
23
24
25