21 Configuration level of the conductivity indicator
ALPH
SiL
SiH
21.5 Configuration parameter for general (not conductivity-
specific) functions
Temperature coefficient [% per °C] of the measured solution.
Value range: 0.00 — 5.50% per °C
The conductivity of a solution varies with the temperature, so for correct
measurement both the temperature and the temperature coefficient of the
solution being measured must be known.
The temperature coefficient can be determined automatically by the
conductivity transmitter, or entered manually.
"Calibrating the temperature coefficient", page 55.
Start of transmission range
This value is taken from the operating instructions for the attached instrument.
Example for the JUMO CTI-Junior, type 202754
(transmission range 0 — 1.00 mS/cm):
SiL = 0.00 mS/cm
End of transmission range
This value is taken from the operating instructions for the attached instrument.
Example for the JUMO CTI-Junior, type 202754
(transmission range 0 — 1.00 mS/cm):
SiL = 1.00 mS/cm