Manual Temperature Setting For Operation Without A Temperature Sensor; Manual Ph Setting For The Rh Measurement Function - JUMO 202710 Operating Instructions Manual

Hand-held meter
Hide thumbs Also See for 202710:
Table of Contents

Advertisement

3.7
Manual temperature setting for operation without a
temperature sensor
3.8

Manual pH setting for the rH measurement function

For precise measurement of pH, mV
the temperature of the medium being measured.
The ideal solution is temperature measurement by a Pt100 temperature sensor
connected to the hand-held meter. The temperature sensor is detected as
soon as it is connected, and its signal is automatically taken into account for
calculating the measured value (ATC: automatic temperature compensation).
If no Pt100 temperature sensor is available, the temperature can be entered by
hand.
Briefly press the
4
min
Use the
or
max
2
5
rH
mV
mV
pH
H
mbar
ATC
ok
CAL
AL Logg
pH
0 to 105°C
or
32 to 221°F
Off
Confirm the entry with the
For exact measurement of rH, the hand-held meter must know the
temperature and pH of the medium being measured. The hand-held meter can
also measure the rH value for a liquid when the temperature and pH values are
entered manually.
Start configuration.
Select the parameter "rH".
Briefly press the
4
The hand-held meter expects a manual temperature entry, see above.
Briefly press the
4
The hand-held meter expects a manual pH entry.
or rH, the hand-held meter must know
H
Set
key.
Menu
key to set the temperature for the measured medium.
max
2
min
5
Setting range for the temperature of the medium.
The manual temperature compensation is switched off.
key.
Store
Quit
6
Set
key.
Menu
Set
key again.
Menu
3 Configuration
13