13.3 Electrode type - C112

C112*
Electrode type
Standard electrode
Special electrode (antimony)

13.4 Electrode monitoring - C114

C114*
Not used
Not used
Not used
Electrode monitoring
Off
On
13 Configuration level of the pH indicator
*The factory-set parameters are shown in the position boxes.
1
* The factory-set parameters are shown in the position boxes.
1
The measurement is monitored for changes. If the
measurement does not change within a defined period, then
it can be assumed that an electrode fault (e.g. glass fracture,
wiring fault, short-circuit) has occurred.
A false alarm may be generated by operational states that
are stationary or change very slowly. Electrode monitoring
should then be switched off.