Leica TPS400 Series User Manual page 46

Hide thumbs Also See for TPS400 Series:
Table of Contents

Advertisement

HzL
α
P0
TPS800_Z158
HZL:
Horizontal angle to a point on the left side of
the object
HZR:
Horizontal angle to a point on the right side
of the object
d:
Distance to the object in the middle between
HZL and HZR
TPS400-5.0.1en
R
P1
d
HzR
Procedure:
CYLINDRICAL OFFSET
Hz Left
Hz Right
V
Hz
PrismOffset:
HzLeft HzRight
1
Enter the prism offset. This is the distance
between the centre of the prism and the surface
of the object to be measured. If the EDM mode
is RL, the value is set to zero automatically.
2
Use the vertical hair, aim at the left side of the
object, then press [HzLeft].
3
Use the vertical hair, aim at the right side of the
object, then press [HzRight].
4
Rotate the instrument accordingly such that the
Hz, the deviation angle, is zero.
46
:
120.4361 g
:
141.4435 g
:
99.4658 g
:
15.398 m
:
0.0000 g
0.030 m
ALL
EXIT
FNC Key