Warnings / Messages
Important Messages
V-Angle not suitable for
calibration (Check V-angle
or face) !
Calibration result out of
tolerance. Previous values
retained !
Hz-Angle out of limit !
Measurement Error. Try
again.
TPS400-5.0.1en
Meaning
Aiming tolerance not met or telescope posi-
tion/face not changed.
Computed values out of tolerance.
Previous values retained.
Hz-angle in second face/telescope pos.
deviates more than 5 gon from the target
point.
Measurement error appeared (e.g. instable
set up or period between measuring in tele-
scope position I and II too long).
Aim on the target point with an accuracy of
min. 5 gon. The target point must be
approximately in the horizontal plane.
Confirmation of the message required.
Repeat measurements.
Confirmation of the message required.
Aim on the target point with an accuracy of
min. 5 gon.
Confirmation of the message required.
Repeat the process.
Confirmation of the message required.