Reading Out The Therapy Data - Atmos S 201 Thorax Operating Instructions Manual

Hide thumbs Also See for S 201 Thorax:
Table of Contents

Advertisement

If the therapy data should be transferred during a patients therapy, follow the steps below:
• Clamp the thoracic catheter
• Stop the current therapy
• Remove the secretion canister.
Perform the therapy data transfer as described.
• Connect the secretion canister.
• Continue the therapy
• Reopen the clamp at the thoracic catheter.

4.7.4 Reading out the therapy data

• Connect the USB flash drive to PC.
• Open the folder on the USB flash drive. This folder contains a PDF file and an excel file.
• Open the PDF file.
• Fill in the desired Information:
-
Patient data
1
-
Diagnosis
2
-
Description of the secretion
3
Following information can be seen in the report:
-
Beginning and end of recording, flow at beginning and end of recording
4
-
File name und device ID
5
-
Graphic diagram of the therapy data
6
Therapy Report
ATMOS® S 201 Thorax V2.0
Patient:
1
Diagnosis:
2
Secretion:
3
Start date:
End date:
16.04.2015 / 08:01
21.04.2015 / 10:13
4
Air leak:
Flow @ Start 0.00 l/min
Flow @ End 0.28 l/min
Complete data transfer
• As soon as the therapy data is transferred then the USB flash
drive may be removed. Now you return to the main screen.
File name:
ATMOS_S_201_THORAX_ID0000_20150421_101400.pdf
Device ID:
0000
Recording time:
00:03:42:00
5
6
27