2.2.3
Changing the DIC prism in the
universal condenser
• To loosen the cover on the condenser underside
(2-25/1), loosen both grub screws using the
SW 2 screwdriver
Position the turret disk containing the DIC prism
to be exchanged in the exchange opening and
hold it on the knurled ring.
• Unscrew retainer ring using the so-called
double-function tool (2-25/4).
• Now screw the tool in the prism mount (2-25/3)
with the threaded side (2-25/4) and pull out DIC
prism (2-25/2).
• Remove DIC prism (2-25/2) from the tool and
screw on the new, required DIC prism.
• Installation of the DIC prism is made in reverse
order. Make sure that the DIC prism is oriented
correctly (groove 2-25/5 must engage in the
pinion of the mount). Make sure that the
knurled ring of the turret disk is labelled
correctly.
2.2.4

Attachment of filter mount

• Hold filter mount (2-26/5) parallel to the
underside of the condenser carrier (2-26/3) and
screw holding pin (2-26/4) of the filter mount
(2-26/5) into the front threaded hole on the left
below the condenser carrier (2-26/3) until stop
using the SW 2 angled Allen key (2-26/6).
• Use adjusting lever (2-26/1) to screw stop bolt
(2-26/2) into the rear threaded hole of the
condenser carrier until stop.
B 40-075 e
10/03
Attachments and conversions
and
remove
the
START-UP
cover.
Fig. 2-25
Fig. 2-26
Changing the DIC prism

Attachment of filter mount

Carl Zeiss
2-17