Attachment Of D Polarizer; Attachment Of Senarmont Dic Polarizer - Zeiss Axioskop 2 plus Operating Manual

Routine microscope
Table of Contents

Advertisement

Carl Zeiss
Fig. 2-27

Attachment of D polarizer

Fig. 2-28
Attachment of SENARMONT DIC
polarizer
2-18
START-UP
Attachments and conversions
2.2.5
• Hold the polarizer (2-27/6)
bottom of the condenser carrier (2-27/3) and
screw in the fastening bolts (2-274) of the
polarizer (2-27/6) with angled adjusting lever
(2-27/5) to the front threaded hole on the left
below the condenser carrier (2-27/3) as far as
the stop.
• Screw in the stop bolt (2-27/2) with the
adjusting lever (2-27/1) as far as the stop in the
rear threaded hole of the condenser carrier
(2-27/3)
2.2.6
• Hold polarizer (2-28/6) parallel to the underside
of the condenser carrier (2-28/3) and screw stop
bolt (2-28/4) of the polarizer (2-28/6) into the
threaded hole on the left below the condenser
carrier (2-28/3) until stop using the angled
adjusting lever (2-28/5).
• Use adjusting lever (2-28/1) to screw stop bolt
(2-28/2) into the rear threaded hole of the
condenser carrier (2-28/3) until stop.
Axioskop 2 mot plus

Attachment of D polarizer

Attachment of SENARMONT DIC
polarizer
B 40-075 e
Axioskop 2 plus
parallel to the
10/03