Do not allow the writing surfaces of the marker insert to get wet or dirty.
Before use, store a refill set (marker insert with cover) in its original packaging at 0 to 25 °C and with
approx. 50 % relative humidity. It is best stored in a refrigerator at approximately 8 °C.
Fig. 2-9

Attaching the universal condenser

• Let the condenser go smoothly; the locking screw will lock in position in front of the groove (2-9/7).
• Screw in centering screws until they engage in the dovetail of the condenser.
Proceed in the same manner when attaching other condensers (e.g. condenser 0.9/1.25 H
with slider for Ph, H, Var).
2-10
START-UP
Initial start-up
2.1.8
Attaching the universal condenser
• Move the stage carrier to the upper stop
position via the focusing drive.
• Use lever (2-9/8) to fold out the front lens on
the condenser. Undo the two centering screws
(2-9/3) until their ends are no longer visible.
• Move condenser carrier (2-9/2) to the lowest
position via drive for height adjustment (2-9/5).
• Insert condenser (2-9/1) between condenser
carrier (2-9/2) and stage carrier (2-9/4). Orient
the locking screw on the bottom of the
condenser in the direction of the groove
(2-9/7).
• Press the dovetail of the condenser against the
spring mount (2-9/6) of the condenser carrier
until the condenser can be placed horizontally
onto the condenser carrier.
Axioskop 40
Axioskop 40 FL
B 40-810 e 12/01