Installing and Removing "Push&Click" Modules
FL P&C reflector module # 000000-1046-281
Optovar module 1,25x (for transmitted-light only) # 000000-1046-284
Optovar module 1,6x (for transmitted-light only) # 000000-1046-283
Optovar module 2,5x (for transmitted-light only) # 000000-1046-282
Analyzer module D # 000000-1050-958
Fig. 2-42
Changing the technique module
Removing a module:
• Tip the module slightly and pull it first out of the top spring elements and then out of the After
removing or installing the reflector modules, insert the cover cap with the top two pins in the stand
housing and press until the bottom retaining elements snap in.
• By switching the reflector turret three positions clockwise, the mounted module swings into the beam.
2-28
START-UP
Attachments and conversions
Installing a module:
• Pull the cover cap (2-42/1) off from the front of
the reflector turret.
• Insert the module (2-42/3) with the retaining
elements (2-42/4) attached to the right and left
of the module from above at an angle into the
spring clamps (2-42/2) on the reflector turret.
• Then press the module at the top until it also
snaps securely into the top spring clamps of the
reflector turret.
Aligning
installing it in the reflector turret:
FL P&C reflector module
Optovar module
Analyzer module
the
respective
the excitation filter is facing forwards;
the emission filter is facing upwards
the lens is facing upwards; the empty
module opening is facing forwards
the analyzer is facing upwards, the empty
module opening is facing forwards
Axioskop 2 plus
Axioskop 2 mot plus
module
when
B 40-075 e