Attachment of active coobservation equipment for 2 users

The active coobservation equipment for 2 users can only be attached to a Axioskop 2 plus /
Axioskop 2 mot plus microscope stand if the incident-light illuminator (e.g. HBO 103) has been removed.
• Use SW 3 ball-headed screwdriver to loosen clamping screw (2-46/12) and remove available tube in an
upward direction.
• Remove dust cap (2-46/14) from the dovetail at (2-46/15) of the coobservation equipment (2-46/16).
• Insert dovetail of coobservation equipment in the stand opening (2-46/13) and tighten clamping screw
(2-46/12) only slightly at first.
• Align coobservation equipment with the outer edges of the stand and make sure that the support
points on the coobservation equipment (below) and the stand (2-46/9) match exactly.
• Then tighten clamping screw (2-46/12) to prevent the coobservation equipment from shifting.
• Unscrew the tube lenses from the two binocular tubes (using the cover of one of the storage cases)
and put them in the cases.
• Remove upper dust caps (2-46/3 and 18) from the coobservation equipment.
• The front dovetail (2-46/17) of the coobservation equipment allows attachment of either the binocular
ergotube 6-25°/23 (2-46/1) or the ergo-phototube 6-25°/23 (100/100). The rear dovetail
(2-46/4) allows not only the attachment of the two ergotubes, but also of the binocular tube 30°/23
(2-46/2) or the binocular phototube 30°/23. Then align the tubes and tighten clamping screws
(2-46/8 and 19).
The binocular ergotube and the ergo-phototube 6-25°/23 provide an upright and unreversed image of
the specimen in the eyepiece.
• Screw heat protection shield (2-46/6) to coobservation equipment from below using two fixation
screws (2-46/7).
• Insert plug (2-46/11) of power unit (2-46/10, 115 V or 230 V) in the socket (2-46/5) on the
coobservation equipment and tighten the screwing ring to prevent shifting.
• Connect power unit to the line (115 V or 230 V).