2.1.7

Attachment of universal condenser

• Move stage carrier to the upper stop position
via the focusing drive.
• Use lever (2-7/8) to fold out the front lens on
the condenser or, in the case of the motorized
version, swing it out by pressing the respective
key. Unscrew both centering screws (2-7/3)
until their ends are no longer visible.
• In the case of the motorized condenser
(Axioskop 2 mot plus), plug connecting cable in
the socket on the left front of the stand (2-8/1)
and tighten screwing ring.
• Move condenser carrier (2-7/2) to the lowest
position via drive for height adjustment (2-7/5).
• Insert condenser (2-7/1) between condenser
carrier (2-7/2) and stage carrier (2-7/4). Orient
the locking screw on the underside of the
condenser in the direction of the groove
(2-7/7).
• Press dovetail of the condenser against the
spring mount (2-7/6) of the condenser carrier
until the condenser can be placed horizontally
on the condenser carrier.
• Let condenser go smoothly, the stud screw will
lock in position at the front of groove (2-7/7).
• Screw in centering screws until they engage in
the dovetail of the condenser.
Other condensers which can be
used
are
achromatic-aplanatic condenser 1.4
H DIC as well as others from the
Axioplan 2 line.
B 40-075 e