b)
Fine adjustment
• Attach illuminator to the stand again and tighten clamping screw (2-11/2).
• Remove diffusion disk (3-1/15) and filters (3-1/18), if switched in, from the beam path.
• Focus on the specimen using objective ≤ 40x and search for a free object area.
• Remove eyepiece and center lamp coil and its reflector image in the pupil image using adjusting
screws (2-12/4 and 5).
• Use adjusting screws (2-12/3) to optimize the homogeneous illumination of the pupil image.
• Switch in diffusion disk and required filters.
4 0
4 0
9 0
9 0
Fig. 2-13
Attachment of HBO 103 illuminator
2-10
START-UP
Initial start-up
2.1.11

HBO 103 illuminator

(1)
Attach HBO 103 W/2 mercury vapor
short-arc lamp
The HBO 103 illuminator and the HBO 103 W/2
mercury vapor short-arc lamp are supplied in
separate packages for safety reasons.
Therefore, insertion of the HBO 103 W/2 lamp into
the lamp housing is the first step to the start-up of
this illuminator.
Insertion of a new HBO 103 W/2 lamp is described
in detail on page 2-19, section 2.2.8 entitled
"Changing the HBO 103 W/2 mercury vapor short-
arc lamp".
(2)
Attachment of HBO 103 illuminator
• Remove cover from the incident-light tube
(2-13/1).
• Insert dovetail of lamp housing in incident-light
tube (2-13/1) on the instrument rear and use
SW 3
ball-headed
clamping screw (2-13/2).
Axioskop 2 plus
Axioskop 2 mot plus
screwdriver
to
tighten
B 40-075 e
10/03