Replacing The Dic Prism On The Universal Condenser - Zeiss Axio Imager 2 Operating Manual

Upright microscope
Table of Contents

Advertisement

ZEISS
Fig. 27
Replacing the DIC prism
• Remove DIC prism (Fig. 27/6) from the tool and screw on the new, desired DIC prism.
• Mount the DIC prism in reverse order, paying particular attention to the correct orientation of the DIC
prism: tongue Fig. 27/5 must be in the recess of the mount in the condenser. Ensure that the labeling
on the knurled ring of the turret disk is correct.
46
FIRST-TIME SET-UP

Replacing the DIC prism on the universal condenser

430000-7544-001
3.15
Replacing the DIC prism on the
universal condenser
Before replacing the prism on the
motorized universal condenser, ensure
that you open the aperture iris first
with the corresponding control button
to prevent damage of the iris blades.
• Use the tool to remove cover (Fig. 27/1), or to
unscrew it (if the cover is at the bottom). Turn
the turret disk until the DIC prism to be
replaced
appears
(Fig. 27/4). Hold the disk by its knurled ring.
• Unscrew the retaining ring using the so-called
dual-function tool (Fig. 27/2).
• Then, screw the tool (Fig. 27/2) with the
threaded side into the prism mount (Fig. 27/3)
and pull out the DIC prism (Fig. 27/6).
Axio Imager 2
in
the
blank
aperture
01/2016