Margin for Negative Absorbance Settings

1 To change the settings select Setup Detector Signals.
2 In the section Margin for Negative Absorbance change the value according
Figure 31
Margin for Negative Absorbance
The higher the value the greater the baseline noise. Set this value only if you expect
N O T E
negative absorbance greater than -100 mAU.

Optimizing the Detector

Additional theoretical information can be found in chapter
the Detector"
Agilent 1260 Infinity DAD and MWD User Manual
to your needs.
on page 99.
Special Settings of the Detector
Use this field to modify the detector's signal handling to increase
the margin for negative absorbance. Use this option if, for
example, your solvent gradient produces a decreasing baseline
absorbance, and for GPC analyses.
Limits: 100 to 4000 mAU.
Using the Detector
"How to optimize
91