Automatic Configuration With Bootp; About Agilent Bootp Service - Agilent Technologies 1260 Infinity G1315C User Manual

Diode array and multiple wavelength detector
Table of Contents

Advertisement

Automatic Configuration with BootP

All examples shown in this chapter will not work in your environment. You need your own
N O T E
IP-, Subnet-Mask- and Gateway addresses.
Assure that the detector configuration switch is set properly. The setting should be either
N O T E
BootP or BootP & Store, see
Assure that the detector connected to the network is powered off.
N O T E
If the Agilent BootP Service program is not already installed on your PC, then install it from
N O T E
your Agilent ChemStation DVD, located in folder BootP.

About Agilent BootP Service

The Agilent BootP Service is used to assign the LAN Interface with an IP
address.
The Agilent BootP Service is provided on the ChemStation DVD. The Agilent
BootP Service is installed on a server or PC on the LAN to provide central
administration of IP addresses for Agilent instruments on a LAN. The BootP
service must be running TCP/IP network protocol and cannot run a DHCP
server.
Agilent 1260 Infinity DAD and MWD User Manual

Automatic Configuration with BootP

Table 29
on page 265.
13
LAN Configuration
273

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents