Optical System

The optical system of the detector is shown in Figure below. Its illumination
source is a combination of a deuterium-arc-discharge lamp for the ultraviolet
(UV) wavelength range and a tungsten lamp for the visible (VIS) and
short-wave near-infrared (SWNIR) wavelength range. The image of the
filament of the tungsten lamp is focused on the discharge aperture of the
deuterium lamp by means of a special rear-access lamp design which allows
both light sources to be optically combined and share a common axis to the
source lens. The achromat (source lens) forms a single, focused beam of light
through the flow cell. Each cell room and lamp are separated by a quartz
window which can be cleaned or replaced. In the spectrograph, light is being
dispersed onto the diode array by a holographic grating. This allows
simultaneous access to all wavelength information.
Figure 1
Agilent 1260 Infinity DAD and MWD User Manual
Optical System of the Detector
1
Introduction

Optical System

11