Detector Error Messages

Detector Error Messages

These errors are detector specific.

Visible Lamp Current

The visible lamp current is missing.
The processor continually monitors the lamp current during operation. If the
current falls below the lower current limit, the error message is generated.
Probable cause
1
2
3
4

Visible Lamp Voltage

The visible lamp voltage is missing.
The processor continually monitors the voltage across the lamp during
operation. If the lamp voltage falls below the lower limit, the error message is
generated.
Probable cause
1
2
134
Lamp disconnected.
Defective visible lamp.
Defective connector or cable.
Defective power supply.
Defective connector or cable.
Defective power supply.
Suggested actions
Ensure the visible lamp connector is seated firmly.
Exchange the visible lamp.
Please contact your Agilent service representative.
Please contact your Agilent service representative.
Suggested actions
Please contact your Agilent service representative.
Please contact your Agilent service representative.
Agilent 1260 Infinity DAD and MWD User Manual