Requirements and Conditions

Parts and Material required
Table 9
analysis. Some of these are optional (not required for the basic system).
Table 9
Agilent 1260
Infinity system
Column:
Standard:
Conditions
A single injection of the isocratic test standard is made under the conditions
given in
Table 10
Flow
Stoptime
Solvent
Temperature
Wavelength
Injection Volume
Agilent 1260 Infinity DAD and MWD User Manual
on page 65 lists the parts and material you need for the set up of the
Parts and Material required
Pump (plus degassing)
Autosampler
Detector, standard flow cell installed
• Agilent ChemStation or
• Instant Pilot G4208 (optional for basic operation) or
with with the appropriate revisions, see
page 24.
System should be correctly set up for LAN communication with the Data
System
Zorbax Eclipse XDB C18, 150 mm x 4.6 mm, 5 µm (993967-906)
Agilent isocratic checkout sample (01080-68704). This 0.5 mL ampoule
contains 0.15 wt.% dimethylphthalate, 0.15 wt.% diethylphthalate, 0.01
wt.% biphenyl, 0.03 wt.% o-terphenyl in methanol.
Table 10
on page 65:
Conditions
1.5 mL/min
8 minutes
100 % (30 % water/70 % Acetonitrile)
Ambient
sample 254 nm (4 nm bandwidth) reference 360 nm (100 nm bandwidth)
1 µL
Using the Detector
Setting up an Analysis
"Performance Specifications"
4
on
65