Installing Capillaries

Installing Capillaries

In May 2013, Agilent has introduced new UHP-FF fittings, which are designed
for improved robustness and ease of use. Previous fittings require careful
handling. Therefore it is important to know, which fittings are used in the
system.
The figure below illustrates the differences between new and previous
capillaries.
Figure 11
New bio-inert capillary and UHP-FF fitting
with nose
For handling instructions of capillaries and fittings, used in modules before delivery of the
N O T E
new UHP-FF fittings (introduced in May 2013), refer to
Cladded PEEK Capillaries"
To work on bio-inert capillaries produced before May 2013, you will need a 1/4 inch
N O T E
wrench instead of the 5 mm mounting tool.
54
Figure 12
Previous bio-inert capillary and fitting
on page 302.
Agilent 1260 Infinity DAD and MWD User Manual