Performance Specifications; Specifications - Agilent Technologies 1260 Infinity G1315C User Manual

Diode array and multiple wavelength detector
Table of Contents

Advertisement

2
Site Requirements and Specifications

Performance Specifications

Performance Specifications

Specifications

Performance Specifications G1315C
Table 3
Type
Detection type
Light source
Data rate
Wavelength range
Short term noise
(ASTM) Single and
Multi-Wavelength
Drift
Linear absorbance
range
Wavelength
accuracy
Wavelength
bunching
Slit width
Diode width
24
Performance Specifications G1315C
Specification
1024-element photodiode array
Deuterium and tungsten lamps
up to 80 Hz
190 – 950 nm
-5
< ± 0.7·10
AU at 254 and 750 nm
-3
< 0.9·10
AU/h at 254 nm
> 2 AU (5 %) at 265 nm
± 1 nm
1 – 400 nm
1, 2, 4 , 8, 16 nm
< 1 nm
Comments
The UV-lamp is equipped with
RFID tag that holds lamp typical
information.
see "Specification Conditions"
below
see "Specification Conditions"
below
see "Specification Conditions"
below
Self-calibration with deuterium
lines, verification with holmium
oxide filter
Programmable in steps of 1 nm
Programmable slit
Agilent 1260 Infinity DAD and MWD User Manual