Controls and Indicators
2.1.13 Manual Release Key Socket <13>
The manual release key can be used as a therapy pause button for all other current
modes. When pressing the release key, the current is switched off immediately. The fol-
lowing symbol appears on the display <3> after pressing the therapy pause button:
You can turn up the current only after setting the intensity control <6> to zero. The ther-
apy time is interrupted as long as the unit is standing by.
2.1.14 VAC Socket <14>
8
The manual release key socket <13> serves to connect a manual
release key for manually triggering the current pulse or as a
therapy pause button. The use of a manual release key may be ap-
propriate when applying the current modes FaS, HVS, T/R, and
KOTS. To activate it, the function display must be at the following
position:
The VAC socket <14> serves to connect to a vacuum application
unit
PHYSIOVAC-Jubilee
treatment, always remove remaining ultrasound gel from the
transducer.
Clean your accessories and instrument on a regular
basis with a disinfecting agent based on aldehyde. By
any means, switch off the device prior to this and pull
the mains plug.
Use a soft sponge cloth for cleaning. Be careful that
no liquid substances invade the instrument. Prior to
cleaning, switch off the instrument and disconnect the
power lead.
Clean and disinfect the sponge and sponge bags after
every treatment, and then rinse them well with fresh
water, so that no residue remains on the material
Regularly check your accessories and replace any
sponges and sponge bags which are contaminated,
damaged or worn out through use, and which are no
longer sufficiently thick.Use a soft sponge cloth for
cleaning. Be careful that no liquid substances invade
the instrument.
Note
The transducer contains a fragile ceramic piezo-resonator! Avoid shoc
the surface and take special care that the transducer is not dropped. Ha
ling malpractice can change the physical characteristics of the resonato
Connecting a Vacuum Unit on page 49).