2.1.12 Ultrasonic Transducer Connector <12>
The ultrasonic transducer is connected to the instrument by means of the ultrasonic
transducer connector <12>. The transducer is automatically calibrated after it has been
connected or when choosing the ultrasound option for the first time after switching on the
instrument.
Replacing the ultrasonic transducer during treatment will not affect the selected mode of
treatment. The dose, however, will automatically be turned down to 0 and has to be in-
creased again when continuing the treatment.
If no transducer has been plugged in, the message
status bar<4>.
Ultrasonic Transducer
Automatic Coupling Confirmation
Both transducer and instrument have an optical coupling confirmation, which guarantees
constant and efficient treatment. If the coupling to the tissue of the patient is not efficient
enough, the light at the transducer will start flashing. You can choose between counting
only the effective treatment time or running the timer independently of the coupling
(Setup Menu).
Warning The transducer contains a fragile ceramic piezo-resonator. Avoid shocks to
the surface and take special care that the transducer is not dropped. Hand-
ling malpractice can change the physical characteristics of the transducer!
IONOSON-Jubilee
Plug in head
The ultrasonic transducer (2.5 cm² and 5 cm²)
works with the frequency 1 MHz. The transducer
is automatically calibrated after connecting or
switching on the instrument.
The transducer is watertight and can be used for
underwater treatment.
Controls and Indicators
will appear in the lower
7