2.1.11 Manual Release Key Socket <11>
The manual release key can be used as a therapy pause button for all other
current modes. When pressing the release key, the current is switched off immedi-
ately. The following symbol appears on the display <3> after pressing the therapy
pause button:
You can turn up the current only after setting the intensity control <6> to zero.
The therapy time is interrupted as long as the unit is standing by.
2.1.12 SIM Socket<12>
2.1.13 VAC Socket <13>
PHYSIOMED-Jubilee
The manual release key socket <11> serves to connect a
manual release key for manually triggering the current pulse or
as a
therapy pause button. The use of a manual release key
may be appropriate when applying the current modes FaS,
HVS, T/R, and KOTS. To activate it, the function display must
be at the following position:
The SIM socket <12> serves to connect to an ultrasound ther-
apy unit
PHYSIOSON-Expert
page 37).
The VAC socket <13> serves to connect to a vacuum applica-
tion unit
PHYSIOVAC-Jubilee
Connecting a Vacuum Unit
Simultaneous Therapy
(see also
or
PHYSIOVAC-Expert
on page 43).
Controls and Indicators
on
(also see
7