Warning Take care that the electrodes are by no means touched after the current is
turned up!
(11) Signs of fatigue should make you stop the treatment (they may already occur after a
few muscle contractions)!
(12) After treatment, an acoustic signal will be issued. Turn the intensity down to 0.
(13) Remove the electrodes.
Treatment with manual release key
Treatment with a manual release key is advisable for individual therapy methods in the
field of sports, for intentional exercises according to Förster as well as for practising both
co-ordination and nerve tracts. This is the great advantage of working with the manual re-
lease key.
(1) Plug the manual release key into the manual release key socket <13> at the rear face
of the instrument.
(2) Apply the electrodes at the respective stimulation points. Position the patient to en-
able free movement of the stimulation area.
(3) Depending on the indication, select one of the current modes
the display <3>.
(4) If necessary, select the function biphase.
(5) Activate the acoustic signal if desired.
(6) If necessary, adapt the pulse time T or the frequency.
(7) When applying FaS or HVS, set the release time R to 1s and the tension time T to the
left to the manual release key position (turn the release time R to the rightmost posi-
tion when applying T/R).
(8) Set the therapy time.
(9) Slowly turn up the intensity with depressed manual release key (intensity control
<6>) until the muscle contraction is sufficient. The therapy time elapses.
Warning When using the manual release key, release a pulse each time you turn up
the intensity!
Warning Take care that the electrodes are by no means touched after the current is
turned up!
(10) Release the button.
If the manual release key is pressed again, the current for FaS and HVS will rise
automatically and will keep the set intensity as long as the manual release key is
pressed. After releasing the button, the intensity will automatically be reduced slowly.
Pressing the button for T/R will trigger a current pulse of the time T previously se-
lected.
(11) Signs of fatigue should stop the treatment (they may already occur after a few mus-
cle contractions)!
(12) After treatment, an acoustic signal will be issued. Turn the intensity down to 0.
(13) Remove the electrodes.
IONOSON-Jubilee
Stimulation Current Therapy
,
or
from
FaS
HVS
T/R