Diagnosis using Stimulation Current

(4) Set the release time R to 1s and the tension time T to the left to the manual release
key position.
(5) Select the ramp of the current rise.
(6) Select the frequency of modulation.
(7) If necessary, alter the basic frequency.
(8) Activate the acoustic signal, if desired.
(9) Set the therapy time.
(10) Slowly turn up the intensity with manual release key pressed (intensity control
<6>) until the muscle contraction is sufficient. The therapy time elapses.
Warning Take care that the electrodes are by no means touched after the current is
turned up!
Warning When using the manual release key, release a pulse each time you turn up
the intensity!
(11) Release the button.
If the manual release key is pressed again, the current at FaS and HVS will increase
automatically and will keep the set intensity as long as the manual release key is
pressed. After releasing the button, the intensity will automatically be reduced in in-
tervals.
(12) Signs of fatigue should stop the treatment (they may already occur after a few mus-
cle contractions)!
(13) After treatment, an acoustic signal will be issued. Turn the intensity down to 0.
(14) Remove the electrodes.
Note
If the intensity (intensity control <6>) is turned up during the time
of release, the following message will appear in the lower status
bar <4>:
Slowly turn down the intensity then. The message will disappear as soon as
the previously set value is reached again.
5 Diagnosis using Stimulation Current
Creating an intensity-time diagram based on IT diagnosis enables you to make statements
on the degree of damage of degenerate muscles as well as suitable parameters for the
therapy.