Controls and Indicators
(LP) Diadynamic Current Modulé en Longues Périodes
acc. to Bernard
Fixed parameters
Pulse shape: sinusoidal half-waves Galvanic basis (5%)
T = pulse time: 10 ms
R = release time: 10 ms
MF basic current with second MF
current alternately every
5 and 10 s.
Galvanic basis ON
(UR)
Ultrastimulation Current
acc. to Träbert
Fixed parameters
Pulse shape: square
T = pulse time: 2 ms
R = release time: 5 ms
Stimulation frequency: about 143
Hz
monophase
T
R
T
Application:
(HV)
High-Voltage Current
Fixed parameters
Pulse shape: twin triangular pulse Mono-/biphase
T = pulse time: 0.1 ms
16
Adjustable
OFF
Adjustable
Mono-/biphase
biphase
T
R
Pains caused by hypertonic muscles,
arthrosis and osteochondrosis.
Adjustable
or adjustable 40 – 400 µs
Stimulation frequency: 2 Hz -
200 Hz (classic: 197 Hz)