Indications Menu
Furthermore, the treatment parameters will be shown. The illustration shows you to
which area of the body you can apply ultrasound. This illustration, however, is only a
suggestion.
(3) Select the function
(4) If necessary, change the treatment parameters.
(5) Increase the ultrasound dose with the dose control <8>. Therapy time elapses.
(6) After treatment, an acoustic signal will be issued. Turn the intensity down to 0.
Automatic Coupling Confirmation
The optical coupling confirmation at both transducer and instrument enables a stable and
efficient therapy. If the ultrasonic waves are badly absorbed by the tissue of the patient,
the light at the transducer will start flashing. Now you can choose between counting the
effective therapy time or running the timer independently from the coupling (setup
menu).
The setup menu provides an additional option to choose whether or not it is possible to
increase the dose after the ultrasonic waves have coupled to the patient's tissue (light at
the transducer goes out).
Warning Make sure you constantly keep moving the transducer on the skin of the
patient (dynamic treatment)! Avoid using the transducer at one spot for a
longer time (static treatment); this form of treatment might lead to damages
on the patient's tissue!
42
.
Therapy
IONOSON-Jubilee