Reprocessing Of Sieve Inserts And Suction Hoses; Preparation Of The Patient Unit; Preparation Of The Tumbler; Preparation Of The Tumbler Holder And Filler - KaVo Primus 1058 Life Care Instructions

Hide thumbs Also See for Primus 1058 Life:
Table of Contents

Advertisement

Care instructions Primus 1058 Life
2 Setup methods according to DIN EN ISO 17664 | 2.6 Preparation of the patient unit

2.5.3 Reprocessing of sieve inserts and suction hoses

Note
Check the sieve inserts daily and after each surgery. The sieve inserts must be re‐
placed at the latest when the suction decreases (Mat. no. 0.763.3423).
Cleaning is not recommended for hygienic reasons.
▶ Remove the suction hoses from the holder.
ð
The suction motor starts automatically when the suction tubes are removed from
the holder.
▶ Hold the cloth under the sieve seat, and pull the sieve insert out of the housing.
▶ If needed, insert a new sieve insert (Mat. no. 07633423) into the seat.
▶ Wipe the suction tubes with disinfectant.

2.6 Preparation of the patient unit

2.6.1 Preparation of the tumbler

▶ Reprocess as specified by the manufacturer.

2.6.2 Preparation of the tumbler holder and filler

Water in the device.
Material damage from leaking water upon actuation of the "Tumbler filler" or "Bowl
flush" keys.
▶ Make sure that the unit is turned off before removing the tumbler filler or the spit‐
toon bowl.
▶ Always check the correct seating of tumbler filler and spittoon bowl before turning
on the unit.
Cleaning and disinfection of the tumbler holder and filler
Note
When assembling, make sure that the O-ring is present on the tumbler filler.
CAUTION
20 / 40

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents