Short instructions for use 6 Hygiene
▶ Switch on the water supply for the treatment unit after the system pressure is
achieved.
▶ Replace the handpiece hoses of the dentist and assistant element in the places
provided in the hygiene centre.
▶ Turn the treatment unit on.
ð The unit automatically starts the remaining steps of the unfinished cycle.
▶ Allow the cycle to complete and then run a second full "weekly" cleaning.
ð The unit is now ready for use.

6.8 Manual internal cleaning of the suction hoses

Note
Clean the suction hoses every evening and disinfect them with DEKA-
SEPTOL Gel daily (included in the evening and weekly cleaning pro-
grams).
▶ Touch all suction hoses consecutively against the extraction point such that no ambi-
ent air is aspirated.
ð An automatically dosed amount of water is supplied by the vacuum. An acoustical
signal is emitted after the process is completed (approx. 3 seconds).
▶ Remove the hose by slowly tipping it to the side, and hook it to the assistant ele-
ment.
36