Reprocessing And Automatic Hygiene Functions Of The Suction System; Reprocessing Of The Suction System Including Suction Hoses - KaVo Primus 1058 Life Care Instructions

Hide thumbs Also See for Primus 1058 Life:
Table of Contents

Advertisement

Care instructions Primus 1058 Life
2 Setup methods according to DIN EN ISO 17664 | 2.9 Reprocessing and automatic hygiene functions of the suction system
▶ Cover the treatment unit during extended periods of non-use.
After the period of non-use
▶ Turn on the central compressor.
▶ Switch on the water supply for the treatment unit after the system pressure is ach‐
ieved.
▶ Replace the instrument hoses of the dentist and assistant units in the places pro‐
vided in the germ reduction attachment.
▶ Turn the treatment unit on.
ð
The unit automatically starts the remaining steps of the unfinished cycle.
▶ Allow the cycle to complete and then carry out a second full intensive germ reduc‐
tion.
ð
The unit is now ready for use.
Additional intensive germ reduction during period of non-use
KaVo recommends carrying out an additional intensive germ reduction at universities
during semester breaks. This flushes the parts of the unit and replaces the germ re‐
duction liquid.
2.9 Reprocessing and automatic hygiene functions of the suction
system

2.9.1 Reprocessing of the suction system including suction hoses

Cleaning and disinfection of the suction system including suction hoses
Note
Clean the suction hoses after each treatment and disinfect them with DEKASEP‐
TOL Gel daily.
Application of DEKASEPTOL Gel
▶ Aspirate one tumbler of cold water with each suction hose.
▶ With the suction handpiece attached or by hand, press down the DEKASEPTOL
dosing facility once each for the saliva ejector hose and the spray mist ejector.
▶ Aspirate the DEKASEPTOL Gel from the dosing facility.
▶ Aspirate another tumbler of cold water with each suction hose.
▶ Finally, aspirate another dose of DEKASEPTOL Gel and allow it to act.
▶ Place the suction hoses in the holder.
DEKASEPTOL Gel basic set
Mat. no. 10007204
DEKASEPTOL Gel refill canister
Mat. no. 10007205