Care instructions Primus 1058 Life
2 Setup methods according to DIN EN ISO 17664 | 2.8 Automatic hygiene functions of the water conducting system
During normal operation of the treatment centre, the continuous germ reduction func‐
tion secures continuous germ reduction in the water conducting systems due to the
automatic dosing of the disinfectant, KaVo OXYGENAL 6.
The rinsing programme and the intensive germ reduction facilitate the automatic rins‐
ing and germ reduction of the water-conducting system. This counteracts the forma‐
tion of micro-organisms in times of stagnation and ensures hygienic water supply for
the patients.

2.8.1 Rinsing programme

The rinsing programme 02 provides for automatic rinsing of the water-conducting sys‐
tem and all the instruments including the tumbler filler for 2 minutes each (hydrogen
peroxide concentration 0.02 %). In the short rinsing programme 01, all instruments on
the dentist unit and assistant unit are rinsed for 20 seconds each.
Carry out the rinsing programme:
▪ before you srat working
▪ after each patient (short rinsing programme)
▶ Briefly press the "Intensive germ reduction" button.
Preparation for rinsing programme and intensive germ reduction
▶ Position the germ reduction attachment on the porcelain spittoon bowl.
Place the instruments on the sterilisation attachment
The cannula of the triple-function and multifunctional handpieces must be removed for
the rinsing program and the intensive germ reduction.
S-table
▶ Lock the swing-arm into the front catching position to ensure that the handpiece is
removed.