Instructions for use
L-MOTOR 181 DB INTRAmatic I - REF 0.535.5060
D D istributed by:
KaVo Dental GmbH
Bismarckring 39
D-88400 Biberach
Phone +49 7351 56-0
Fax +49 7351 56-1488

Table of contents

T T able of contents
1
User instructions ....................................................................................................................................................................... 7
2
Safety ................................................................................................................................................................................... 10
2.1
Description of safety instructions ..................................................................................................................................... 10
2.1.1
Description of safety instructions: Warning symbol ................................................................................................ 10
2.1.2
Description of safety instructions: Structure .......................................................................................................... 11
2.1.3
Description of safety instructions: Description of danger levels ................................................................................. 12
2.2
Safety instructions ........................................................................................................................................................ 14
Table of contents
3 3
Product description .................................................................................................................................................................. 19
3.1
Purpose - Intended use ................................................................................................................................................ 20
3.2
Technical Data ............................................................................................................................................................. 22
3.3
Transportation and storage conditions .............................................................................................................................. 24
3.4
Operating conditions ..................................................................................................................................................... 26
4
First use ................................................................................................................................................................................ 27
4.1
Check the pressure ....................................................................................................................................................... 28
5
Operation .............................................................................................................................................................................. 30
5.1
Connecting the drive motor to the supply hose ................................................................................................................... 31
Manufacturer:
Kaltenbach & Voigt GmbH
Bismarckring 39
D-88400 Biberach
www.kavo.com
1
2

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the L-MOTOR 181 DB INTRAmatic I and is the answer not in the manual?

Questions and answers

Summary of Contents for KaVo L-MOTOR 181 DB INTRAmatic I

  • Page 1: Table Of Contents

    Instructions for use L-MOTOR 181 DB INTRAmatic I - REF 0.535.5060 D D istributed by: Manufacturer: KaVo Dental GmbH Kaltenbach & Voigt GmbH Bismarckring 39 Bismarckring 39 D-88400 Biberach D-88400 Biberach Phone +49 7351 56-0 www.kavo.com Fax +49 7351 56-1488...
  • Page 2 Disinfection: Automated external and internal disinfection ..................49 Drying ..................................49 Care products and systems - Servicing ..........................50 7.5.1 Care products and systems - Servicing: Care with KaVo Spray ................52 7.5.2 Care products and systems - Servicing: Care with KaVo SPRAYrotor ............... 53 7.5.3 Care products and systems - Servicing: Care with KaVo QUATTROcare ..............
  • Page 3: User Instructions

    User instructions 1 1 User instructions Dear User Congratulations on purchasing this KaVo quality product. By following the instructions below you will be able to work smoothly, economically and safely. © Copyright by KaVo Dental GmbH User instructions S S ymbols...
  • Page 4: Description Of Safety Instructions: Structure

    Safety 2 2 .1.2 Description of safety instructions: Structure DANGER The introduction describes the type and source of the hazard. This section describes the potential consequences of non-observance. ▶ The optional step includes necessary measures for hazard prevention. Safety 2 2 .1.3 Description of safety instructions: Description of danger levels The safety instructions listed here, together with the three levels of danger will help avert property damage and injury.
  • Page 5 To ensure proper function, the medical device must be set up according to the reprocessing meth‐ ods described in the KaVo Instructions for Use, and the care products and care systems descri‐ bed therein must be used. KaVo recommends specifying a service interval at the dental office for a licensed shop to clean, service and check the functioning of the medical device.
  • Page 6: Product Description

    Product description 3 3 Product description L-MOTOR 181 DB INTRAmatic ( Mat. no. 0.535.5060 Product description 3 3 .1 Purpose – Intended use Purpose: The motor is: ▪ an air motor intended to operate/drive a dental handpiece or contra-angle handpiece ▪...
  • Page 7: Transportation And Storage Conditions

    Product description Air consumption up to 60 Nl/min Idle torque 2.3 Ncm at 2.2 bar (32 psi) The INTRAmatic L motor can be attached to all hoses with Borden 2-hole connection. With INTRAmatic 1 coupling. All straight and contra-angle handpieces with INTRAmatic connection ISO 3964 can be attached. Product description 3 3 .3 Transportation and storage conditions CAUTION...
  • Page 8: First Use

    First use 4 4 First use WARNING Hazard from contaminated products. Infection danger to the care provider and patient. ▶ Before first use and after each use, prepare the product. First use 4 4 .1 Check the pressure CAUTION Compressed air connections on device. Dirty and moist compressed air causes premature bearing wear.
  • Page 9: Connecting The Drive Motor To The Supply Hose

    Operation 5 5 .1 Connecting the drive motor to the supply hose ▶ Attach plug-in fitting to the motor connection and screw tight in clockwise direction using the sleeve. Note Details on the hose connection can be obtained from the manufacturer. Operation 5 5 .2 Regulating the spray water CAUTION...
  • Page 10: Removing The Attachment Instrument From The Drive Motor

    Operation ▶ Attach the KaVo instrument to the motor until it locks in place. ▶ Pull on it to make sure that the instrument is securely attached to the motor. Operation 5 5 .4 Removing the attachment instrument from the drive motor ▶...
  • Page 11: Troubleshooting

    ▶ Slide the O-rings forwards and remove them. Troubleshooting ▶ Insert new O-rings into the grooves and spray them with KaVo spray. ▶ Press the O-rings between your fingers to form a loop. ▶ Slide the O-rings forwards and remove them.
  • Page 12: Cleaning

    Preparation methods according to ISO 17664 ▶ The medical device must be dry when transported for reconditioning. ▶ Do not place it in a solution or similar. ▶ Recondition the medical device as soon as possible after treatment. 7 7 .2 Cleaning CAUTION Malfunction of the product from cleaning in the ultrasonic unit.
  • Page 13: Disinfection: Manual External Disinfection

    Preparation methods according to ISO 17664 7 7 .3.1 Disinfection: Manual external disinfection KaVo recommends the following products based on material compatibility. The microbiological ef‐ ficacy must be ensured by the disinfectant manufacturer. ▪ Mikrozid AF Liquid made by Schülke & Mayr ▪...
  • Page 14: Care Products And Systems - Servicing: Care With Kavo Spray

    Note KaVo only guarantees that its products will function properly when the care products used are those listed as accessories, as they were tested for proper use on our products. Preparation methods according to ISO 17664 7 7 .5.1 Care products and systems - Servicing: Care with KaVo Spray...
  • Page 15: Sterilisation

    Preparation methods according to ISO 17664 KaVo recommends servicing the product after each time it is used, i.e. after each cleaning. ▶ Servicing the product. See also: Instructions for Use K K aVo QUATTROcare Preparation methods according to ISO 17664 7 7 .6 Sterilisation...
  • Page 16: Warranty Terms And Conditions

    Warranty terms and conditions KaVo shall not be liable for defects and their consequences that have arisen or may arise from natural wear, improper handling, cleaning or maintenance, non-compliance with operating, main‐ tenance or connection instructions, calcination or corrosion, contaminated air or water supplies or chemical or electrical factors deemed abnormal or impermissible in accordance with KaVo's in‐...