2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit
Sterilisation of S-table handle and holder pad/anti-slip mat
Damage caused by sterilisation.
▶ Do not sterilise the white standard handle.
▶ Sterilise the holder pad/ anti-slip mat.
The grey handle (Mat. no. 1.002.4489) can be sterilised (see symbol on the handle).

2.4.2 Reprocessing of the instrument holder of the TM table

Note
The handpiece holder (Mat. no. 3.001.4229) can be removed to facilitate cleaning.
Cleaning and disinfection of the instrument holder
Damage caused by thermodisinfection.
▶ Do not thermodisinfect the instrument holder.
▶ Place the instrument hoses in the provided places of the germ reduction attach‐
ment during the cleaning and germ reduction of the handpiece holder.