Care instructions Primus 1058 Life
3 Replenish and check the disinfectant | 3.1 Replenish OXYGENAL 6
▶ Quit replenishing immediately if you hear a signal sound being issued every sec‐
ond (10 sounds).
ð
Reservoir is full.
▶ Attach the tumbler holder and the tumbler filler.
Note
Check the O-ring on the tumbler filler for damage.

3.1.1 Replenish OXYGENAL 6 when using the water bottle

Overdosing OXYGENAL.
▶ Using the water bottle in combination with water block DVGW, there is no need to
add more OXYGENAL 6 into the water bottle.
Damage caused by defective water bottle.
Malfunctions or failures from a leaky water bottle.
▶ Do not wash the water bottle in a dish washer, and do not rinse it with water hot‐
ter than 50 ℃.
▶ Do not use the water bottle if it is scratched, deformed or discoloured.
▶ Note the maximum useful life of the water bottle on the label.
Water bottle complete: (Mat. no. 0.710.4151)
Dosing attachment: (Mat. no. 1.002.0287)
CAUTION
CAUTION
34 / 40