3 Replenish and check the disinfectant | 3.2 Check the OXYGENAL 6 concentration
Note
Substances remaining in the dosing head are not contamination, but rather agents
that are deposited there during extended periods of non-use after condensation of
the water. These deposits redissolve after several dosing processes.
▶ Screw the water bottle tightly to the treatment centre in a clockwise rotation.
ð
Apply compressed air to the water bottle.

3.2 Check the OXYGENAL 6 concentration

▶ Fill the tumbler with water.
▶ Use Merckoquant peroxide test strips 100 to 1000 mg/l H
0.229.5003) to determine the OXYGENAL 6 concentration in the water germ re‐
duction system. The OXYGENAL 6 concentration should be in the range from 150
to 250 mg/l H
O
in normal operation.
2
2
37 / 40
Care instructions Primus 1058 Life